In its default configuration, the affected product transmits plain-text patient data to a hard-coded public IP address when a patient is hooked up to the monitor. This could lead to a leakage of confidential patient data to any device with that IP address or an attacker in a machine-in-the-middle scenario.
CVE ID: CVE-2025-0683
CVSS Base Severity: HIGH
CVSS Base Score: 8.2
CVSS Vector: CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:N/VA:N/SC:N/SI:N/SA:N
Vendor: Contec Health
Product: CMS8000 Patient Monitor
EPSS Score: 0.04% (probability of being exploited)
EPSS Percentile: 11.77% (scored less or equal to compared to others)
EPSS Date: 2025-02-28 (when was this score calculated)